CL543B thermocouple and RTD calibrator, thermocouple and RTD calibrator (PT100), calibrate thermocouple instruments to 0.1 and 0.01, °F and °C
[CL543B Thermocouple and RTD Calibrator — Features]
Thermocouple and RTD calibrator (PT100)
Calibrate thermocouple instruments to 0.1 and 0.01, °F and °C
Types J, K, T, E, R, S, B, N, G, C, D, L (J-DIN), (T-DIN), and P (Platinel): -13.000 to 80.000 mV
Calibrate RTD instruments to 0.1 and 0.01, °F and °C
Pt 10, 50, 100, 200, 500, and 1000 (3850)
Pt 100 (alpha = 3902, 3916, 3926)
Cu 10 and 50 as well as Ni 120
Calibrate with confidence
Resolution of 0.01, °F and °C; accuracy to 0.2°
Cold junction sensor accurate to 0.05°C
NIST-traceable cold junction sensor in isothermal block
Troubleshoot RTD sensors and instruments
Auto-detects 2-wire, 3-wire, or 4-wire connections; indicates when individual wire connection is open
Verify correct instrument operation by measuring excitation current
Quick and easy calibration
Instantly select 3 outputs using EZ-CHECK switch
Easily find trip points using 2-speed EZ-DIAL knob and auto-step
Verify resistance, cold junction temperature, and output
Secondary display of resistance for RTD and compensated mV settings, as well as cold junction temperature for thermocouples
For factory, field, or shop
Includes rubber boot and mV test leads
Slim switches prevent accidental damage
Brass screw and mini thermocouple socket for thermocouple wires
Perform heat treatment surveys and accuracy tests
Meets or exceeds AMS 2750 requirements as a secondary standard instrument and field test instrument
NIST certificate (no points)
Thermocouple Connection
Simulating or reading thermocouples requires the use of thermocouple or extension-grade thermocouple wire. Insert the thermocouple wire into the mini thermocouple jack or place the bare thermocouple wire under the brass block below the screw.
The CL543B thermocouple and RTD calibrator has two banana jacks (1+ and 2-) mounted on top of the housing. These are not temperature compensated and are for mV signal only.
RTD (PT100) Connection
Simulating or reading RTDs uses copper wire. Insert 2-wire, 3-wire, or 4-wire into the corresponding jacks on the calibrator. For RTD source, the CL543B simulates the (+) RTD from jacks 1 and 4 and the (−) RTD from jacks 2 and 3. When reading an RTD sensor, the CL543B thermocouple and RTD calibrator automatically detects whether it is a 2, 3, or 4-wire connection using proprietary circuitry. This aids in troubleshooting sensor connections.