DP9601 high-precision digital RTD thermometer
Product Introduction:
DP9601 high-precision digital RTD thermometer/data logger, reference standard thermometer for Pt100 sensors, Omega USA temperature calibrator- Consultation Hotline:18615230752 Landline:0531-88975421
Product Details
Technical parameters
Application scenarios
Optional accessories
Contact Us
DP9601 high-precision digital RTD thermometer/data logger, reference standard thermometer for Pt100 sensors, Omega USA temperature calibrator
[DP9601 High-Precision Digital RTD Thermometer/Data Logger — Features]
High-precision digital RTD thermometer/data logger
Reference standard thermometer for Pt100 sensors
High precision: ±0.01°C Pt100
40 millikelvin (0.040°C) system accuracy
Simple, fast digital-matched calibration of sensors; up to 10 points per channel retained in memory
Alpha-numeric display
0.01°C resolution for Pt100 input
Data logging
Wide temperature range
Readings directly in °C, °F, Kelvin, or ohms
Two sensor input ports
Channel A, B, or A-B readings
USB serial communication
USB flash drive port
Includes PC software
Programmable analog retransmission output
Rechargeable lithium-ion battery / mains powered
The DP9601 high-precision digital RTD thermometer/data logger provides exceptional platinum resistance thermometer temperature measurement and logging for demanding laboratory and field applications. It is an extremely accurate and stable digital thermometer, particularly suitable for use with thermal calibration baths as a reference standard. Applications include medical, pharmaceutical, food, environmental testing, R&D, and educational institutions. Displayed values and user information appear on a bright LED screen with diffused backlight. Data can be displayed in °C, °F, K, and ohms as desired; nine front panel keys provide intuitive user controls used in conjunction with the display. This arrangement makes operation extremely simple and “user-friendly.” There are two input ports for 3-wire or 4-wire Pt100 sensors; the instrument automatically recognizes 3-wire or 4-wire configuration. Measured temperatures can be displayed directly from one input, or differential readings can be displayed between the two inputs. Differential temperature and both individual channel temperatures are displayed simultaneously. A flash drive USB port is provided to allow data storage and/or export. PC software running in Windows® is standard; it allows programmable custom calibration, remote control, and measurement and recording functions.
[DP9601 High-Precision Digital RTD Thermometer/Data Logger — Specifications]
Inputs/Ranges/Sensors
Type: Pt100 to IEC 751 (reference ITS 90) -200 to 850°C (-328 to 1562°F), Ro = 100Ω; 3-wire or 4-wire connection with automatic recognition (with manual operation)
Overall Accuracy: ±0.01°C ±0.0005% of span
Linear Conformity: Better than ±0.01°C
Stability (vs. ambient temperature): Better than 0.0025°C per 1°C ambient change
Warm-up: Negligible under normal ambient conditions; 5 to 10 minutes required for full stabilization unless stored at low temperature, then at least 30 minutes
Pt100 Sensor Current: 0.5 mA
Data Display Resolution: 0.01
Variable Filter: Sampling rate selectable from 4 to 64 (average measurements per reading)
Measurement Units: °C, °F, K, Ω
Measurement Mode: A, B, or A-B
Custom Calibration: Up to 10 calibration values can be assigned to Pt100 inputs A and B. Values are retained in non-volatile memory until replaced by user; ITS 90 or IPTS 68 coefficients can be used for custom calibration
Smart Sensor Connection: Correction values stored in the connector
Null Function: Corrects differential temperature reading between two sensors to zero
Sensor Lead Resistance: Maximum 5Ω per lead
Recording: 8000 readings
Power Supply: Built-in lithium-ion rechargeable battery; mains adapter includes universal 90 to 260V 50/60 Hz; battery life up to 12 hours on full charge
Series Mode Rejection: 60 dB @ 50 Hz (applied 50 mV RMS)
Common Mode Rejection: Applying 30V RMS between input and ground produces no measurable effect
Display: 128 x 64 pixel character OLED screen with diffused backlight; wide viewing angle, high contrast (adjustable)
Front Panel Controls: 9 membrane keys to control all instrument functions
Mechanical/Housing: Metal desktop housing / adjustable tilt
Dimensions: 190 W × 70 H × 250 mm D (7.5 × 2.75 × 9.85")
Weight: Approximately 3 kg (6.61 lb)
Input Connections: 2 × Pt100 via D-type connector
USB Serial: Isolated, 38400 baud, 8 data, no parity, 1 stop bit; remote control and measurement
PC Software (standard): Provided as standard on CD-ROM; remote control and measurement; record readings to file / download to PC / program corrections
Analog Output (standard): User-programmable, 0 to 1 Vdc
Application Note: Inputs are not isolated within the instrument; primarily intended for laboratory applications and field calibration of industrial temperature sensors. Therefore, probes connected to the instrument must be isolated from high voltage pickups.
Precision Pt100 Probe: 304 stainless steel probe, 6 mm diameter; with 2 m shielded PTFE lead wire, Pt100 probe and connector
DP9600-L250: 250 mm (9.85") L, -50 to 250°C (-58 to 482°F)
DP9600-H450: 350 mm (13.75") L, -50 to 450°C (-58 to 842°F)

186-1523-0752